Liver/Lung
首页> Product > Clinical Application

1.Placing position of electrode needle 

Inserting position of umbrella electrode:

                                                             


When the first point of the ablation of the umbrella electrode  is ended, the needle shall move back 1cm to ablate the second point until the entire tumor is completely ablated. 

 

2.Clinical Operation Guidelines: The parameter setting (power  control) of generator for tumor RF ablation 
Electrode with expanded array diameter of 2cm:

1) Long time, simple operation

Type

Control Mode

Expected ablation size

Initial power

Operation

Maximal power

Diameter of 2cm

Power control

Ellipsoidal, with major axis of 2.0cm-2.5cm

30W

Adjust power between initial power and maximal power during operation according to temperature feedback and keep it between 90-110℃ at best. Ablation time should be no less than 5 min.

80W

1 Power can be adjusted flexibly during operation. Generally, you can adjust it to a higher power if temperature increases slowly, vise versa.

2 Ablation size is proportional to power multiplied by ablation time.

 

2)Short time, slightly complex operation

Type of electrode

Control mode

Expected ablation range

Setting power

Operation in the treatment

diameter of 2cm

Power control

Ellipsoidal, with major axis of 2.0cm-2.3cm

70W or 80W

When the ablation is completed, and the machine cuts off automatically RF output, the RF sub needle shall be taken back, and open again after rotated a certain angle in situ, and carry out ablation again.

Electrode with expanded array diameter of 3cm:

1) Long time, simple operation

Type

Control Mode

Expected ablation size

Initial power

Operation

Maximal power

Diameter of 3cm

Power control

Ellipsoidal, with major axis of 3.0cm-3.7cm

50W

Adjust power between initial power and maximal power during operation according to temperature feedback and keep it between 90-110℃ at best. Ablation time should be no less than 8 min.

110W

1 Power can be adjusted flexibly during operation. Generally, you can adjust it to a higher power if temperature increases slowly, vise versa.

2 Ablation size is proportional to power multiplied by ablation time.

 

2)Short time, slightly complex operation

Type of electrode

Control mode

Expected ablation range

Setting power

Operation in the treatment

diameter of 3cm

Powercontrol

Ellipsoidal, with major axis of 3.0cm-3.3cm

100W or 110W

When the ablation is completed, and the machine cuts off automatically RF output, the RF sub needle shall be taken back, and open again after rotated a certain angle in situ, and carry out ablation again.

 


 

Electrode with expanded array diameter of 4cm:

1) Long time, simple operation

Type

Control Mode

Expected ablation size

Initial power

Operation

Maximal power

Diameter of 4cm

Power control

Ellipsoidal, with major axis of 4.0cm-4.7cm

60W

Adjust power between initial power and maximal power during operation according to temperature feedback and keep it between 90-110℃ at best. Ablation time should be no less than 10 min.

130W

1 Power can be adjusted flexibly during operation. Generally, you can adjust it to a higher power if temperature increases slowly, vise versa.

2 Ablation size is proportional to power multiplied by ablation time.

 

2)Short time, slightly complex operation

Type of electrode

Control mode

Expected ablation range

Setting power

Operation in the treatment

diameter of 4cm

Power control

Ellipsoidal, with major axis of 4.0cm-4.5cm

120W or 130W

When the ablation is completed, and the machine cuts off automatically RF output, the RF sub needle shall be taken back, and open again after rotated a certain angle in situ, and carry out ablation again.

Electrode with expanded array diameter of 5cm:

1) Long time, simple operation

Type

Control Mode

Expected ablation size

Initial power

Operation

Maximal power

Diameter of 5cm

Power control

Ellipsoidal, with major axis of 5.0cm-6.0cm

70W

Adjust power between initial power and maximal power during operation according to temperature feedback and keep it between 90-110℃ at best. Ablation time should be no less than 15 min.

150W

1 Power can be adjusted flexibly during operation. Generally, you can adjust it to a higher power if temperature increases slowly, vise versa.

2 Ablation size is proportional to power multiplied by ablation time.

 

2) Short time, slightly complex operation

 

Type of electrode

Control mode

Expected ablation range

Setting power

Operation in the treatment

diameter of 5cm

Power control

Ellipsoidal, with major axis of 5.0cm-5.5cm

140W or 150W

When the ablation is completed, and the machine cuts off automatically RF output, the RF sub needle shall be taken back, and open again after rotated a certain angle in situ, and carry out ablation again.


 


免责声明

欢迎访问迈德医疗官方网站(以下简称“本网站”)。在开始使用前,请您务必仔细阅读并完全理解以下免责声明。您一旦访问或使用本网站,即视为您已接受本声明的全部条款。

1. 用户身份界定

本网站旨在为医疗卫生专业人员提供相关参考信息与服务。如果您非前述专业人员,请立即停止使用本网站。您应对使用本网站的行为自行承担全部责任。

2. 非专业建议

本网站所载所有信息,包括但不限于文本、数据、图像等,仅供参考之用,绝不意图替代医疗卫生专业人员的独立判断和医学决策。在做出任何与健康相关的决策或行动前,您应咨询合格的医疗卫生专业人员。

3. 知识产权

本网站包含的所有内容,其知识产权(包括但不限于版权、商标权、专利权)为迈德医所有或有权使用。未经迈德医疗事先书面许可,任何人均不得复制、修改、分发或用于任何商业目的。

4. 国际性与产品可用性

本网站信息具有地域属性。并非所有在本网站提及的产品、服务、适应症或注册信息均在您所在的国家或地区提供。相关产品的具体供应情况和相关参考信息,请咨询您所在地的迈德医疗销售代表或授权代理商。

5. 不保证与责任限制

迈德医疗致力于提供准确和及时的信息,但对本网站及其内容不作任何明示或默示的保证,包括但不限于对准确性、完整性、及时性、适销性、特定用途适用性及不侵犯他人权利的保证。 在任何情况下,迈德医疗及其关联方、董事、员工均不对因使用或无法使用本网站、或因依赖其内容所作决策而导致的任何直接、间接、附带、惩罚性、特殊或结果性损失或损害(包括但不限于数据丢失、利润损失、业务中断、设备损坏)承担任何法律责任。

6. 用户责任与病毒风险

您承诺以合法方式使用本网站,不得从事任何损害网站安全或干扰他人使用的行为。迈德医疗对因访问、浏览、使用本网站或下载任何内容而可能导致的计算机病毒、恶意代码等对您计算机系统造成的损害或损失不承担责任。您应自行采取充分的防病毒措施。

7. 第三方链接

本网站可能包含指向第三方网站的链接。这些链接仅为提供便利,并不意味着迈德医疗对其内容的认可或保证。迈德医疗对任何第三方网站的内容、隐私政策或服务不承担任何责任。

8. 前瞻性陈述

本网站可能包含基于当前预期和假设的前瞻性陈述。这些陈述本质上存在已知和未知的风险与不确定性,实际结果可能会与前瞻性陈述中的预测存在重大差异。迈德医疗不承担更新这些前瞻性陈述的义务。

9. 服务变更与终止

迈德医疗保留在不事先通知的情况下,随时修改、暂停或终止本网站(或其任何部分)服务的权利。

10. 管辖法律与可分性

本声明的解释、适用及争议解决,均适用中华人民共和国法律。因本声明或使用本网站所产生的任何争议,均应提交至迈德医疗注册地上海市浦东新区人民法院诉讼解决。若本声明任何条款被认定为无效,不影响其余条款的持续有效性和执行力。